9. Operating Instructions – Change Program
9-10 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
WARNING
Your Fill Volume in milliliters (mL) should normally not exceed the
values shown in Table 19-7 on page 19-15. Contact your dialysis
center to confirm your Fill Volume if it exceeds this volume. Exceeding
this volume can result in an increased intraperitoneal volume (IIPV)
situation.
IIPV could result in a feeling of abdominal discomfort, serious injury,
or death. See page 18-58 for IIPV symptoms or if IIPV is suspected.
Additional care should be taken to monitor for IIPV symptoms for
those patients not able to communicate essential information to their
caregiver during treatment.
Fill Volume
Volume of solution for each nighttime cycle based on your
prescription.
Minimum setting is 100 mL
Maximum setting is 3000 mL
Default setting is 250 mL
Setting increments:
10 in the 100 to 500 mL range
50 in the 500 to 1000 mL range
100 in the 1000 to 3000 mL range
Table 9-3. CCPD/IPD Settings – Standard Fill Mode (Continued)
Setting Description
FILL VOL: ML