Red Cycler 5C Betriebsanleitung Seite 129

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 377
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 128
07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-35
9. Operating Instructions – Change Program
9.8.1 CCPD/IPD Therapy Settings
(Low Fill Mode)
Table 9-11 shows the settings programmed for CCPD/IPD therapy in the Low Fill
Mode.
Table 9-11. CCPD/IPD Settings – Low Fill Mode
Setting Description
Total Volume
Total Volume of solution used for the therapy. Includes the
total Fill Volume for all cycles and the Last Fill Volume.
Minimum setting is 200 mL
Maximum setting is 80000 mL
Default setting is 200 mL
Setting increments:
50 in the 200 to 2000 mL range
100 in the 2000 to 20000 mL range
500 in the 20000 to 80000 mL range
Therapy Time
Total time for the nighttime portion of therapy. This time is
fixed and starts with the Initial Drain.
Minimum setting is 10 minutes
Maximum setting is 48 hours
Default setting is 10 minutes
Setting increment is 10 minutes
THERAPY: CCPD/IPD
TOTAL VOL: ML
THERAPY TIME: HH:MM
Seitenansicht 128
1 2 ... 124 125 126 127 128 129 130 131 132 133 134 ... 376 377

Kommentare zu diesen Handbüchern

Keine Kommentare