18. Troubleshooting
18-6 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
To Correct
Cause 1:
1. Check the line referenced on the display screen for:
Kinks
Closed clamps
Fibrin blockage
Disconnected solution bags
Empty solution bags
2. Correct the issue you found.
You do not need to press any buttons.
– OR –
If the alarm becomes a Manual Restart Alarm:
1. Press
STOP to mute the alarm.
2. Check the line referenced on the display screen for:
Kinks
Closed clamps
Fibrin blockage
Disconnected solution bags
Empty solution bags
3. Correct the issue you found, if possible.
– OR –
If the issue is not corrected, contact your dialysis center.
4. Press
GO to return to the therapy.
CHECK LINE (Continued)