07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 11-15
11. Operating Instructions – Prepare for Therapy
11.7 Connect the Solution Bags
h
NOTE:
Connect yourself only when CONNECT YOURSELF appears on the
display screen.
WARNING
If the solution bag on the heater pan does not cover the
heater button on the right end of the heater, delivery of
overheated solution can result. In some circumstances, this
can cause injury to the patient. Additional care should be
taken when positioning small solution bags on the heater
pan.
WARNING
If you use a Last Fill with a different solution, the line with
the BLUE clamp must be connected to the Last Fill solution
bag. If the solution concentration is incorrect or the wrong
solution is connected to the Last Fill line, UF alarms can
occur during therapy because the wrong solution can
generate too much or not enough ultrafiltrate volume.
If you discover that your therapy was performed with an
incorrect solution, or an incorrect solution was connected
to the Last Fill (BLUE clamp) line, contact your dialysis
center.
WARNING
Follow aseptic technique taught by your dialysis center
when handling lines and solution bags to reduce the
possibility of infection. Always put on a face mask and wash
and dry (or disinfect) your hands thoroughly.