07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-57
18. Troubleshooting
18.7 Reprime Patient Line Procedure
If the fluid level is not at or near the connector at the end of the patient line, follow
the steps below to reprime the patient line.
Steps to reprime the patient line Display screen
1. Press
STOP when the display screen alternates
between CONNECT YOURSELF and
CHECK PATIENT LINE.
alternates with
2. Press V until REPRIME PATIENT LINE appears.
3. Press
ENTER.
4. Verify that the patient line is properly primed:
a. Make sure fluid is present near the connector
at the end of the patient line.
The display screen alternates between
CONNECT YOURSELF and CHECK PATIENT LINE.
5. Repeat Steps 1 through 4 until the patient line is
primed.
CONNECT YOURSELF
CHECK PATIENT LINE
REPRIME PATIENT LINE
a