07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-19
18. Troubleshooting
18.3.8 Slow Flow Rate
SLOW FLOW
Display Message:
Cause: Flow rate is very slow. A slow flow rate can reduce the Dwell
Time and decrease the amount of effective dialysis time.
There can be a partial kink on the specified line.
This is an Auto Restart Alarm.
To Correct: 1. Check the line shown in the display for:
– Kinks
– Closed clamps
– Fibrin blockage
– Disconnected solution bags
– Empty solution bags
2. Correct the problem you found.
You do not need to press any buttons.
– OR –
If the alarm becomes a Manual Restart Alarm:
1. Press
STOP to mute the alarm.
2. Check the line shown in the display for:
– Kinks
– Closed clamps
– Fibrin blockage
– Disconnected solution bags
– Empty solution bags
3. Correct the problem you found.
4. Press
GO to return to the therapy.