18. Troubleshooting
18-34 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
The HomeChoice APD System usually recovers from the System Error. Device
swaps are necessary only if the System Error repeats with different disposables.
SYSTEM ERROR nnnn
Message Display:
Cause: A problem occurred inside the system in the current cycle or
over several cycles.
To Correct: 1. Press
STOP to mute the alarm.
2. Write down the System Error number and therapy
phase that appears on the display screen.
3. For instructions and assistance, contact Baxter
Technical Assistance at the number listed in
2.2, Numbers to Call for Assistance, on page 2-1.
4. Follow the instructions from Baxter Technical
Assistance. Your next steps will vary, as shown in
Table 18-2.
Table 18-2. Steps to Correct System Error Alarms
System Error Steps to Correct
2042, 2044, 2046 If this alarm occurs before you are connected, treat as a
Reload Set Alarm. 18.3.5, Load a New Set, on page 18-12.
Otherwise treat as a System Error.
2065 to 2071,
2098, 2265
Close all clamps.
Press the power switch Off and On to end the therapy.
2240, 2267, 2367 See 18.3.16, System Errors 2240 or 2267, on page 18-31 for
these alarms.
All others One of the following:
Continue the therapy
End the therapy early
Bypass a phase
Perform a Manual Drain
SYSTEM ERROR nnnn