07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-5
9. Operating Instructions – Change Program
Basic steps to change settings Display screen
1. Press
V to change or review settings (before you
press
GO to start your therapy).
– OR –
Press
V during your therapy.
2. Press
ENTER to access the CHANGE PROGRAM
menu.
THERAPY is the first setting that appears.
3. If you do not want to change this setting, press
V
to see TOTAL VOL (Total Volume).
4. Press
ENTER to change the setting, if needed.
(The option or value blinks)
5. Press U and V to change the value.
6. Press
ENTER to save the new value.
(Blinking stops)
7. Press V to display THERAPY TIME.
NITE THER TIME (Night Therapy Time) appears
for Hi-Dose therapies.
8. Make changes, if needed, by following Steps 4–6.
9. Press
V to display FILL VOL (Fill Volume).
NITE FILL VOL (Night Fill Volume) appears for
Hi-Dose therapies.
10. Make changes, if needed, by following Steps 4–6.
PRESS GO TO START
CHANGE PROGRAM
THERAPY: CCPD/IPD
TOTAL VOL: 15000ML
TOTAL VOL: 15000ML
TOTAL VOL: 14000ML
TOTAL VOL: 14000ML
THERAPY TIME: 8:00
NITE THER TIME: 8:00
FILL VOL: ML
NITE FILL VOL: ML