18. Troubleshooting
18-16 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
LOW DRAIN VOLUME
Display Message:
Cause:
A Slow Flow, No Flow, or air and fluid condition
occurred before the programmed Minimum Drain
Volume % (or Initial Drain Volume) completed.
A No Flow condition occurred before the Low Fill Mode
Minimum Drain Time completed.
This is an Auto Restart Alarm.
To Correct: 1. Change your position to try to drain more fluid.
2. Check for kinks in your patient line.
3. Correct the problem you found.
You do not need to press any buttons.
– OR –
If the alarm becomes a Manual Restart Alarm:
1. Press
STOP to mute the alarm.
2. Change your position or lower the cycler by six (6)
inches.
3. Press
GO to return to your therapy.
If the alarm continues during your therapy, or if the alarm
occurs regularly over multiple therapies:
1. Follow the steps below to check your Drain Volume.
2. Contact your dialysis center to learn when it is safe to
bypass. See 18.4.5, Bypass Low Drain Volume Alarm, on
page 18-45.